2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 595
P- A.0.0 ISOLATE POWER CORD / PLUG / STRAIN RELIEF
a) Unplug the machine!
b) Inspect the power plug and cord for burning or other damage.
c) Per the Figure below, inspect the strain relief for damage.
d) Plug into the known good GFI outlet.
(page 595).
Figure 112 – Heater Breaker Switch / Main Power Switch / Strain Relief
P- A.2.0 ISOLATE HEATER CIRCUIT
A)
IMPORTANT!
Per the Figure above, turn the
Heater Breaker Switch
OFF!
B) ENSURE the Main Power Switch REMAINS ON!
C) Press the Power button. If the machine STAYS ON perform parts D and E. If the machine DOES NOT
TURN ON (no fan) OR it TURNS ON but immediately ‘blows’ the GFI skip to part F.
D) Return to Dialysis OR preferably Heat Disinfect.
E) Allow up to thirty (30) minutes, for the problem to possibly reoccur, BEFORE continuing to part F.
F) Per the ORIGINAL symptom, FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF was ‘blowing’ the GFI but DOES NOT now:
(page 596).
2)
IF continues to ‘blow’ the GFI:
Proceed to
page 649
, procedure number
3)
IF was NEVER turning on (fan NEVER runs) AND still NEVER turns on:
Proceed to
page
, procedure number
4) IF was NEVER turning on but is on now (fan running):
The heater
1
may be bad. This can be
confirmed by turning the machine off, turning the Heater Switch on, and attempting to turn the
machine on again.
1
To LOCATE the Heater refer to
(page 129).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...