2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 503
SECTION 14 - UF (ULTRAFILTRATION) PUMP PROBLEMS
UF- 1.0.0 ISOLATE UF CIRCUIT
THREE (3) checks! Answer for all three below Check #3:
Check #1:
Does the “UF Pump Alarm” banner appear?
Check #2:
Call debug screen 0. Figure right, is the UF Pump ‘dot’ ALWAYS blue?
(i.e. NEVER white)
Check #3:
Figure right, is the UF Pump symbol PINK?
Yes (to any of the three):
See procedure number UF- 5.0.0 (
page 508
No (to ALL three):
Perform parts a AND b below:
a) Per the Figures below AND right, TWO more checks:
Check #4:
ENSURE the UF Pump is installed with
its output, red or blue, nozzle at the TOP AND is
mated to the mounting plate’s output
(→) arrow.
Check #5:
ENSURE the OUTPUT tubing is
connected to UF Check Valve #63; INPUT tubing, at
the white nozzle, to UF Filter #74.
b) If no problems were located above see procedure number
(page 504).
Figure 85 – Hydraulics Top / Front Views
The UF Pump ‘dot’ toggles between white
and blue when the UF Pump strokes. It
should NOT be constantly blue!