2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 194
TSV- 2.0.0 ISOLATE VALVE WIRES
a) THREE (3) checks, WHILE tracing the
NOTED
valve’s wires from
the distribution board to the solenoid terminals (Figure right):
Check #1:
ENSURE no pinched or bare wires!
Check #2:
ENSURE both wires terminate PROPERLY at the
valve’s solenoid terminals!
Check #3:
ENSURE no corrosion at the solenoid terminals!
b) If no problems were located, see procedure number
(page 194).
TSV- 2.1.0 ISOLATE VALVE SOLENOID
a) Figure right, open the cap from the NOTED valve’s
distribution board connector.
b)
Except for Valve #26
, the wires MUST BE
connected between the top and bottom terminals.
Valve #26 (ONLY!) MUST BE connected between the
second from top and bottom terminals! If wired
incorrectly this is the problem!
c)
Set your CALIBRATED volt
meter to RESISTANCE (Ω)!
d) Where the wires are connected, place one meter lead on one terminal and the other lead on the other.
e) Figure right, reading the meter’s numeric AND
UNITS (Ω), between 40 and 100
Ω
[Yes or No]?
Yes
Between 40 and 100 Ω! FIVE (5) possible bad components (see Component List below).
Swap in one at a time and in between test the machine.
Component List: 1)
The
NOTED
valve;
2)
Actuator-Test Board;
3)
Actuator cable;
4)
Functional Board*;
5)
Distribution board
* To prevent “Cond Offset Failure” place the machine into T and C (refer to
OPERATING MODES
(page
Error! Bookmark not defined.
))
No
Less than 40 Ω OR more than 100 Ω! See procedure number
(page 194).
TSV- 3.0.0 CHECK SOLENOID
Figure right, place one meter lead on one solenoid terminal and the other lead
on the other terminal. Between 40 and 100
Ώ
?
Yes
Between
40 and 100 Ω! The valve’s wiring harness is bad.
No
Less than 40 Ω OR more than 100 Ω! The valve’s solenoid is bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...