2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 291
T- 4.0.4 ISOLATE NTC #2 CONNECTIONS
a)
Turn the machine OFF!!
b) Per the Figure right, place the
274 Ω
plug, from the
TWO-RESISTOR
SET
, into the
1
st
distribution board
position from the left, “CON-NTC”.
c)
Set your volt meter to measure resistance (Ω)!
d)
ENSURE the black lead remains attached to chassis ground (see
page 4)!
below, measure at the solder side of SENSOR BOARD’S ‘X2’ connector at
pin 13
(bottom
row, 7 pins from the rear of machine).
Figure 51 – NTC #2 Checks
f)
Figure right, reading the meter’s numeric AND
units display, TWO (2) possible scenarios:
1)
IF (and ONLY if) LESS THAN 300
Ω (0.300 KΩ):
ENSURING the machine was OFF before
measuring AND assuming all procedures were performed correctly the mother board may be bad.
2) IF more than 300
Ω:
See parts a THROUGH d below:
a) ENSURE the machine was OFF prior to the measurement!
b) ENSURE the
274 Ω
plug, from the
TWO-RESISTOR SET
, is placed properly in the
1
st
distribution board positon from the left, “CON-NTC”! If not, repeat procedure number
c) ENSURE the meter was set to measure resistance (
Ω
) AND its go=round lead is at chassis! If
(page 291).
d) TWO (2) possible bad components: 1) Sensor Board
*
cable OR; 2) Distribution board. *
NOTE:
The Sensor Board cable can be checked!
NOTE
that one (1)
NTC #2
connection will be checked
and proceed to
page 524
SECTION 17- CHECKING THE SENSOR BOARD CABLE
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...