2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 184
CLEAN- 5.0.0 ISOLATE POSSIBLE VALVE ELECTRICAL PROBLEM
Leaving the shunt door OPEN:
a) Call debug screen 1 (Figure right).
If debug does not appear press
‘Escape’ then call screen 0.
b) Does
VERR
(lower right column)
= 0?
Yes
VERR
= 0
!
Proceed to
page
, procedure number
No
VERR
= 1 OR more! See procedure number
(page 184).
CLEAN- 5.0.1 VERR DOES NOT = 0 / ISOLATE SPECIAL VALVE ERROR
a) Call debug screen 0
(Figure right).
b) Ignoring the TOP
window,
Flow Error,
in part c you will
watch the
2
nd
window down,
Valve Error
!
c) Ignoring a ‘blink to 1’ that lasts less than one (1) second, watch for one (1) minute. Does
Valve Error
EVER = 1 for LONGER THAN two (2) seconds CAREFUL HERE! TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve Error = 0 OR ‘blinks to 1’ for less than one (1) second!
Close the
shunt door then
proceed to
page 206
TROUBLESHOOTING VALVE ERRORS IN CLEANING
2)
IF Valve Error EVER = 1 LONGER THAN two (2) seconds:
Turn the machine OFF then
proceed to
page 189
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...