2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 230
T- 1.0.17 SIMULATE CONDUCTIVITY
a)
Figure right, leaving the 6.04 KΩ resistor in, plug the
274 Ω
resistor plug, from the
TWO-RESISTOR
SET
into the Cond Cell’s distribution board position,
“X7, COND”.
b) Watch the [
Conductivity
] window for one (1) minute.
THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) BETWEEN 12.9 and
14.5 mS AND does NOT change more
than 0.2:
(page 230).
2)
IF (and ONLY if) BETWEEN 12.9 and 14.5 BUT changes MORE than 0.2:
THREE
(3) possible bad components (see COMPONENT LIST below). Turn the machine off and swap in
each, one at a time, in between returning to Dialysis Program, until [
Conductivity
] is STABLE
between 12.9 and 14.5 mS.
COMPONENT LIST: 1)
Sensor Board
a
;
2)
Power Logic Board;
3)
Functional Board
a
.
a
To prevent “Cond Offset Failure”
place the machine in
T and C Mode
(refer to
OPERATING
MODES
, page
Error! Bookmark not defined.
))
3)
IF less than 12.9 OR more than 14.5 mS:
See parts a AND b below:
a) Remove the
274 Ω
plug then reinsert it OR consider using the
274 Ω
plug from a different
TWO-RESISTOR SET.
b) If (and ONLY if) [
Conductivity]
is still is less than 12.9 OR more than 14.5proceed to
page
, procedure number
. If NOW STABLE between 12.9 and 14.5 see procedure
(page 230).
T- 1.0.18 CONDUCTIVITY IS BETWEEN 12.9 AND 14.5 / VERIFY CONDUCTIVITY
Based on the [
Conductivity
] window’s color, TWO (2) possible scenarios:
1)
IF (and ONLY if) pale / yellow white:
(page 231).
2)
IF RED:
See parts a THROUGH d below:
a) At the bottom of the screen, select the ‘Dialysate’ tab.
b) Figure right, if necessary adjust the Limits until
the ‘Actual’ Conductivity is CENTERED between
them.
c)
IMPORTANT!
Press ‘CONFIRM’.
d) Allow two (2) minutes or
until if the
[Conductivity]
window turns white. TWO (2) possible scenarios
next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...