2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 259
T- 1.7.72 TEMP = 4.0 OR MORE
a)
Turn the HEATER Switch OFF!
b)
Return NTC #2 to the 1
st
distribution board position from the left, “CON-NTC”!
c) Temperature may continue to increase for a few minutes! Allow the external meter to fall below 35.0° C
BEFORE continuing to procedure number
(page 259).
T- 1.7.8 VERIFY TEMPERATURE
a)
Turn the HEATER Switch ON!
b) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
d) Allow five (5) minutes BEFORE continuing to part e!
e) Based on the Home screen’s [
Temperature
] reading. TWO (2) possible scenarios:
1) IF LESS THAN 35.1° C:
(page 260).
2) IF BETWEEN 35.1 and 38.9° C:
A Temperature problem is not present at THIS time. To check
possible intermittent temperature problems proceed to
page 222
, procedure number
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...