2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 134
MOTORS- 4.1.0 MOTOR ROTATING?
Is the
NOTED
motor rotating now (Yes or No)?
Yes
Motor is rotating! If (and ONLY if) a bad connection was corrected above AND if a Flow Error
occurs see the Table of Contents for whatever Program the machine is in! If a bad connection
was NOT located above proceed to
page 138
, procedure number
No Motor is NOT rotating! See procedure number
(page 134).
MOTORS 4.2.0 ISOLATE PROGRAM
What Program is the machine CURRENTLY in
1)
Rinse OR
2)
Dialysis?
1)
IF (and ONLY if) in RINSE:
Proceed to
page 135,
2)
IF in Dialysis Program:
(page 134).
MOTORS- 4.3.0 IN DIALYSIS PROGRAM / ISOLATE EMPTYING PROGRAM
Call debug screen 2 to see
! EMPTY
(left column). TWO (2) possible scenarios:
1)
IF (and ONLY if) ! EMPTY = 0:
Either the (red) arterial dialyzer connector is NOT connected to the
shunt properly OR FOUR (4) possible bad components*;
1)
Bad Actuator-Test Board;
2)
Bad
Functional Board** OR;
3)
Bad arterial line shunt door micro switch OR;
4)
Bad motherboard
2)
IF ! EMPTY = 1:
See parts a AND b below:
a) Open the shunt door and remove both dialyzer connectors!
b) CLOSE THE DOOR. From debug screen 2, look at
! EMPTY
again! TWO (2) possible scenarios:
1)
IF (and ONLY if) ! EMPTY = 1:
A)
Return the dialyzer connectors to the shunt.
B)
2)
IF ! EMPTY = 0:
FOUR (4) possible bad components*:
1)
Bad Actuator-Test board;
2)
Bad Functional Board** OR;
3)
Bad shunt door switches OR;
4)
Bad motherboard.
*
Swap in each component one at a time, with known good, until
! Empty
= 1
** To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...