2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 463
TMP- 3.1.0 ACFS BETWEEN 3.5 AND 6.0 / ISOLATE DIALYSATE PRESSURE TRANSDUCER
This procedure uses the CFS transducer (known good) to check the dialysate pressure transducer.
a) Per the Figure right, unplug the Dialysate Pressure
Transducer #9 from distribution board position “x9,
PDIAL”.
b) Check inside the vacant x9 position for corrosion
and/or damaged male pins.
Damage indicates
the distribution board may need to be
replaced!
c) Per the Figure right, CAREFULLY place the CFS
Transducer’s distribution board connector (#10) into the Dialysate Pressure Transducer’s position “x9,
“PDIAL”.
d) From debug screen 10, is
PDIA
(middle column) between 3.5 and 6.0?
Yes
PDIA
between 3.5 and 6.0! Dialysate Pressure Transducer #9 is bad.
No
PDIA
is NOT between 3.5 and 6.0! See parts A through E below:
A) Leave CFS (#10) plugged into “X9, PDIAL” for parts B through E.
B) Turn the machine OFF!
C) One at a time, swap in the listed components, (see
COMPONENT LIST
below), with
known good and in between, until
PDIA
is between 3.5 and 6.0!
COMPONENT LIST: 1)
Sensor Board
1, 2
;
2)
Actuator-Test Board
1
;
3)
Sensor Board cable;
4)
Functional Board
1, 2
;
5)
Distribution board;
6)
Motherboard.
1
To LOCATE the boards, refer to
(page 9)
2
To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
)
D) Return to Dialysis Program.
E) If screen 10’s
PDIA
is now between 3.5 and 6.0 the last component swapped in was the
probem. If
PDIA
is NOT between 3.5 and 6.0 return to part B
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...