2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 123
F- 16.0.7 LESS THAN 130 ML PER MINUTE COLLECTED / ISOLATE INCOMING WATER
a) Figure right, place the
274 Ω
resistor plug, from the
TWO-RESISTOR SET
, into the float’s distribution
board position, “X5, FLOAT-SW”.
b) Allow forty (40) seconds or until flow from the Vent
Tubing occurs indicating hydrochambers A
through C are full!
c) MEASURE flow from the Vent Tubing for ONE (1) minute. More than eight hundred (800) ml collected?
Yes
More than 800 ml per minute! See procedure number
(page 123).
No Less than 800 ml! Call debug screen 0. Figure right, is Valve
#41’s ‘dot’ BLUE?
Yes
‘Dot’= Blue!
A)
Reattach the Deaeration Pump’s input tubing;
B)
Proceed to
page 141
No ‘Dot’ = White’! See parts a AND b below
a) BE VERY SURE the
274 Ω
resistor plug, from the
TWO-RESISTOR SET
, is
placed PROPERLY at distribution board position “FLOAT-SW”! If not, repeat
(ABOVE) procedure number
(page 123).
b) FIVE (5) possible bad components:
1)
Actuator-Test Board OR;
2)
Sensor
Board cable OR
3)
Sensor Board OR;
4)
Distribution board OR;
5)
Motherboard.
F- 16.0.8 MORE THAN 800 ML / ISOLATE AIR LOCK
a) Figure right, avoiding the VACANT position on the left, return
the float’s connector to distribution board position, “X5,
FLOAT-SW”!
b) Attach the
60 ml syringe
, with its plunger pushed all the
way in, to the Deaeration Pump’s clear (INPUT) tubing.
c) PULL on the syringe plunger to the end of the barrell! If Restrictor #48 is still open you should feel very
little resistance!
d) Remove the syringe and MEASURE flow from the tubing for one (1) minute. If less than 130 ml repeat
parts b through d. After no less than
ten (10) attempts
do you MEASURE more than 130 ml every
minute?
Yes More than 130 ml every minute! See procedure number
(page 124).
No Cannot achieve more than 130 ml per minute! THREE (3) possible bad components:
1)
Reoccurring debris at Restrictor #48 OR;
2)
Bad Float Switch* OR;
3)
Bad hydrochamber.
* To LOCATE the Float Switch refer to
(page 129)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...