I) Watching for thirty (30) seconds to ENSURE a “No Water” or Flow Error NEVER occurs, is gauge
pressure cycling to a consistent
peak
of between 18 and 20 psi?
Yes
Peak between 18 and 20 psi! See procedure number
(page 666).
No NOT between 18 and 20 psi! Referring to the Figure (previous page), loosen Regulator’s
#61’s lock nut and turn the center screw until a consistent peak of between 18 and 20 psi is
achieved. TWO (2) possible scenarios:
1)
IF (and ONLY if) pressure adjusts to between 18 and 20 psi:
See procedure number
(page 666).
2)
IF pressure will NOT adjust to between 18 and 20 psi:
TWO (2) possibilities:
1)
Incoming water pressure is more than 105 psi OR; 2
)
Regulator #61 is bad. Regulator
#61 can be rebuilt (Rebuild kit, P/N 190934).
NOTE:
Previous high pressure may have
been already damaged the Heat Exchanger causing it to leak!
HEATEX- 2.0.0 ISOLATE INLET PRESSURE REGULATOR #61
This procedure checks that Inlet Pressure Regulator #61 maintains pressure over time:
a)
IMPORTANT!
Leaving the water
ON
, turn the machine
OFF
.
b) Watch the gauge for two (2) minutes. Pressure should not increase to more than 20 psi! If it does
Regulator #61 is bad but can be rebuilt (Rebuild kit, P/N 190934). If it does NOT increase to more than
20 psi NOTE this pressure then see part c.
NOTE:
Previous high pressure may have been already damaged the Heat Exchanger causing it to leak!
c) Allow twenty (20) minutes. Does pressure increase more than 2 psi above what was noted in part b?
Yes
Pressure increases! TWO (2) possibilities:
1)
Incoming water pressure is more than 105 psi OR;
2)
Bad Regulator #61. Regulator #61 can be rebuilt (Rebuild kit, P/N 190934).
NOTE:
The Heat
Exchanger may have already been damaged causing it to leak!
No Pressure does NOT increase! Per the Figure below, repair or replace the Heat Exchanger.
Figure 136 – Heat Exchanger (Exploded)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...