2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 83
Yes
BC Switch
goes t
o
897 or more, even if only once!
Read before performing!
Return to
page 26
, procedure number
but, if you return here, because of no drain flow,
THREE (3) possible bad components (see COMPONENT LIST below):
COMPONENT LIST: 1)
Actuator-Test Board
1
OR;
2)
Actuator cable OR;
3)
Multiple
intermittent restricted balancing chamber valves #32, #34, #36 and #38
2
.
1
To LOCATE the Actuator-Test board refer to
(page 9);
2
To LOCATE these valves refer to
(page 84)
No
BC Switch
IS NEVER, EVER 897 or more.
A Flow Error is not presenting at this time! It may be
extremely intermittent. Return to
page 44
, procedure number
LEFT BLANK INTENTIONALLY