2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 460
TMP- 2.3.2 POSITIVE PRESSURE TEST
a) CLAMP the external meter’s tubing segment!
The external meter is invalid during the next test!
b) SLOWLY PUSH on the syringe plunger until screen 0’s
PDial
= between positive (+)300 and (+)325!
WARNING!
DO NOT 325 mmHg!
c) Clamp the Four-Way Assemblies ‘to syringe’ tubing segment. Does
PDial
HOLD +/- 25 per minute
(Yes or No)?
Yes
PDial
HOLDS +/- 25 mmHg per minute! See parts a AND b below:
a)
IMPORTANT!
Return the dialyzer lines to the shunt and CLOSE THE DOOR!
b) Proceed to
page 461
, procedure number
No
PDial
does NOT hold! See procedure number
(page 460).
TMP- 2.3.3 PDIAL DOES NOT HOLD / ISOLATE POSITIVE PRESSURE LEAK
a)
IMPORTANT!
Return the dialyzer lines to the shunt AND CLOSE THE DOOR!
b) A
procedure, in a different Section, is performed next.
NOTE
this page and procedure number
(TMP- 2.3.3) because you may prompted to return here.
c) BEFORE continuing to part d, proceed to
page 496
, to perform
d) If a leak was not located in part c, check the Four Way for a positive pressure leak!
e) Place the machine in HEAT DISINFECT and allow Temperature to increase to more than 70° C!
f) CAREFULLY check the hydraulic compartment, the ENTIRE LENGTH of the dialyzer lines, and the
DiaSafe
®
housing for external leaks.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...