2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 203
VE- 6.0.0 VALVE ERROR AT BALANCING CHAMBER CYCLE 2
Valves #31, #34, #36, and #37 ‘dots’ REMAINING blue. One of these valve
circuits is causing the problem and these procedures isolate between them.
a) At the distribution board (Figure below), unplug
Valve #31’s
connector.
b) Call debug screen 0 and watch the Balancing Chamber Valve ‘dots’ for
forty-five (45) seconds. Do they begin to cycle between white and blue?
Yes The Balancing Chamber Valves begin to cycle between white
and blue! Valve #31 OR its blue wire harness is bad.
No Valves DO NOT cycle! See procedure number
(page 203).
Figure 39 – Valve Cycle 2
VE- 6.0.1 VALVE #31 UNPLUGGED AND VALVES DO NOT CYCLE
a) At the distribution board (Figure above), unplug
Valve #34’s
connector.
b) Watch the Balancing Chamber Valve ‘dots’ for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Yes The Balancing Chamber Valves cycle between white and blue! Valve #34 OR its blue wire
harness is bad.
No Valves DO NOT cycle! See procedure number
(page 203).
VE- 6.0.2 VALVE #34 UNPLUGGED AND VALVES DO NOT CYCLE
a) At the distribution board (Figure above), unplug
Valve #36’s
connector.
b) Watch the Balancing Chamber Valve ‘dots’ for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...