2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 88
F- 7.8.0 PARTS 3 AND 4 YIELDED 800 ML OR MORE / RETURN SYSTEMS
All secondary side components are checking good HOWEVER, the problem may be intermittent.
a) Turn the machine OFF!
b) Reattach the drain tubing. ENSURING it is properly attached to the station!
c) Return Valves #31 AND #38 to their distribution board positions.
d) Return Flow Pump #21 AND Deaeration Pump #20 to their distribution board positions.
e)
OPEN THE SHUNT DOOR till instructed otherwise
(page 88).
F- 7.8.1 CONFIRM SYSTEMS
a) Turn the machine ON but
DO NOT
press any keys yet!
b) From “Select Program”, if the pumps are plugged in correctly, the Deaeration motor rotates but the Flow
motor DOES NOT!
c) Re
turn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
d)
SLAM
the Loading Pressure gauge in the Acid/Acetate Rinse port! ENSURE PEAK pressure is
between 23 and 27 psi.
e) See
procedure number
(page 88).
F- 7.8.2 TROUBLESHOOT FLOW ERROR
a) ENSURE Dialysate Flow is ON (Dialysate Flow lamp IS NOT blinking).
b) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
c) ENSURE the FLOW motor is rotating i.e. plugged in properly!
d) ENSURE a “No Water” alarm NEVER occurs.
e) Call debug screen 6 to watch
BC Switch
for three (3) FULL minutes OR until if it ever becomes more
than 400 OR less than 203 indicating a flow problem!
TWO (2) possible scenarios below:
1)
IF (and ONLY if) BC Switch is
NEVER
more than 400 OR less than 203:
A)
Leaving the shunt
door open, allow Temperature and Conductivity to become normal;
B)
Return to (ABOVE)
page 44
2) IF BC Switch is or becomes less more than 400 OR less than 203 even if only once:
See parts
a AND b below:
a) CLOSE THE SHUNT DOOR!
b) Return to (ABOVE) procedure number
page 30
).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...