2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 174
Yes
Between 34 and 36 psi! Proceed to
page 175
, procedure number
No Remains less than 34 psi! ENSURING Heat Disinfect OR Rinse is running AND no leaks, see
parts a THROUGH c below:
a) Watch for one (1) minute to ENSURE a “No Water” alarm NEVER occurs!
(page 158), behind the Deair Pump, locate
Valve #78
.
c) Can you adjust Valve #78 until the gauge cycles to a PEAK of between 34 and 36 psi?
Yes
Pressure between 34 and 36 psi! Allow two (2) minutes then, if a Flow Error
reoccurs, proceed to
page 175
, procedure number
No Pressure DOES NOT adjust. TWO (2) possible scenarios:
1)
IF (and ONLY if) pressure remains low:
See procedure number
(page 174).
2) IF pressure remains high:
Valve #78 may be bad!
CLEAN- 3.2.2 LOW FLOW PUMP PRESSURE / ISOLATE VALVE #43
a)
Per the Figures below
, DOUBLE clamp the OUTPUT tubing at Valve #43.
NOTE:
Valve #43’s
output tubing extends towards the front of the machine!
b) Can you now adjust Valve #78 until gauge pressure cycles to between 34 psi and 36 psi about every
three (3) seconds (Yes of No)?
Yes
Between 34 and 36 psi! This may indicate Valve #43 sticking open!
A)
Release the clamps!
B)
ENSURING Valve #43’s ‘dot’ is white, if pressure again DOES NOT cycle to more than
34 psi then TWO (2) possible bad components:
1)
Bad Actuator-Test Board OR;
2
) Bad
Valve #43.
No Remains less than 34 psi! Leaving the clamps in place, proceed to
page 94
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...