2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 251
T- 1.5.5 SCREEN [TEMPERATURE] = 95° C
A) Turn the machine OFF!
B) Figure right, insert the 80° C
(1.255 KΩ) plug, from the
FOUR-
RESISTOR SET
, into the
2
nd
distribution board position from the
LEFT, “MON-NTC”. This simulates
approximately 80° C.
C) Return to HEAT DISINFECT with an external temperature
meter attached as described previously!
D) IMMEDIATELY call debug screen 10. If debug does not appear press ‘Esc’ then call screen 10.
E) Watch
5V HI
AND
24V HI
for three (3) minutes OR until if
5V HI
= 5.3 or more and / or
24V HI
= 28.0 or
more?
Yes
5V HI
exceeds 5.3 and / or
24V HI
exceeds 28.0! Proceed to
page 253
, procedure number
No
5V [HI]
remains less than 5.3 AND
24V HI
remains less than 28.0! Continue to parts F through H.
F) To prevent damage DO NOT leave the machine! The external meter MUST NEVER exceed 95° C!
G) Press ‘Esc’ to call the main Heat Disinfect screen.
H) Watch the screen’s [
Temperature
] window for one (1) minute. THREE (3) possible scenarios:
1)
IF (and ONLY if) screen [Temperature] is between 77 and 83° C AND stable i.e. does NOT
change more than 1.0:
Leaving the resistor plug installed, proceed to
page 255
, procedure
2)
IF (and ONLY if) screen [Temperature] is unstable i.e. changes more than 1.0:
Leaving the
resistor plug installed, proceed to
page 254
, procedure number
3)
IF screen [Temperature] is stable i.e. DOES NOT change more than 1.0 but IS NOT between
77 and 83° C:
As prompted perform parts A THROUGH L below:
A) ENSURE t
he 80° C (1.255 KΩ) plug, from the FOUR-RESISTOR SET, was placed properly at
the
2
nd
distribution board position from the left, “MON-NTC”! If not, repeat procedure number
(page 251).
B) Consider using the
80° C (1.255 KΩ) plug, from another FOUR-RESISTOR SET. If screen
[
Temperature
] is still NOT between 77 and 83° C continue to part C.
C) Remove the resistor plug and return NTC #3’s connector to the
2
nd
distribution board position
from the left, “MON-NTC”.
D) BEFORE continuing to part E (next page), place the machine into RINSE for fifteen (15) minutes
to cool it down rapidly.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...