2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 496
SECTION 12 – INDUCED POSITIVE PRESSURE TESTS
During these procedures hydraulic systems are checked for external leaks.
USE A FLASHLIGHT! BE
THOROUGH!
FEEL AROUND!
Small leaks may take several minutes to become obvious and cause BIG
problems!
1.
Place the machine into RINSE!
This creates positive pressure in the secondary circuit!
2. Figure below check for leaks at:
1)
Top and bottom of Air Removal Chamber #69;
2)
Filter #74;
3)
UF
Pump #22. If no leaks continue to step #3.
Figure 80 – Internal Dialysate Circuit Leak Checks
3. Figure below, check for leaks at:
4)
Valve #78;
5)
Dialyzer Quick Connectors and O-rings;
6)
External
Flow Indicator #75;
7)
Dialysate Line Filter #73;
8)
(Optional) Fluid Sample Port;
9)
Dialysate line
connections at the rear of the machine. If no leaks continue to step #4 (next page).
Figure 81 – External Dialysate Circuit Leak Tests
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...