
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 410
FIL- 1.0.0 ISOLATE AIR SENSOR CIRCUIT / POSSIBLE LEAKS
Going forward, if a “No Water”, Flow Error, Temperature or a Conductivity EVER occur address them first!
a) Figure below, remove the Distribution Board’s cover.
b) Using a flashlight, ENSURE the Air Sensor’s female connector is plugged PROPERLY
*
into its
Distribution Board position i.e. 4
th
connector, 5
th
position, from the LEFT!
*
IMPORTANT!
If CBE modified the
CBE board
plugs directly into the distribution board and positions
the female Air Sensor connector two (2) pins higher than the others! The female
connector MUST plug into the CBE board pin for pin!
*
IMPORTANT!
Using a flashlight,
ENSURE
the TOP CBE board pin is covered by the Air Sensor’s
connector! If not,
FILACT
will always = 1!
NOTE:
If NOT CBE modified do so immediately as this may solve Filling Program problems. For
questions call 1-800-227-2572 and reference Field Action FA2014-01..
Parts c through e next page
Figure 65 – Chamber #69 / Air Sensor
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...