2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 648
P- 3.0.1 REMAINS ON FOR AT LEAST 30 SECONDS
With the machine on, QUICKLY unplug the
sixteen (16) pin, Switch Matrix Cable from the Front Panel
Interface board. TWO (2) possible scenarios:
1)
IF the machine stays on after unplugging the cable:
Place the machine into Heat
Disinfect and with the shunt door open allow up to one (1) hour. If the machine STAYS on see
procedure number
(page 648). If it turns itself off again see OTHER POSSIBILITIES below.
OTHER POSSIBILITIES:
Swap in the following components (see Component List below) one at a
time, with known good, and in between if the machine remains on the last component swapped in
is the problem.
Component List:
1)
Power Logic Board;
2)
Power Logic Board cable;
3)
Main Relay (K1)*.
*
To LOCATE K1 refer to
(page 651).
2)
IF the machine turns off before unplugging the cable:
The most likely cause is the front
panel (Power button). Replace the front panel HOWEVER, if the problem reoccurs see OTHER
POSSIBILITIES below:
OTHER POSSIBILITIES:
Swap in the following components (see Component List below) one at a
time, with known good, and in between, if the machine remains on the last component swapped in
was the problem:
Component List:
1)
Power Logic Board;
2)
Power Logic Board cable;
3)
Main Relay (K1)*.
*
To LOCATE K1 refer to
(page 651).
P- 3.0.2 MACHINE REMAINS ON
The most likely culprit is the front panel (Power button) HOWEVER, this can be confirmed by, while leaving
the machine on unplug the
sixteen (16) pin, Switch Matrix Cable from the Front Panel Interface board
Allowing up to one (1) hour, if the machine turns itself off again the front panel is bad!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...