2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 594
7) IF the machine NEVER turns on i.e. fan NEVER runs:
Perform parts a THROUGH d below:
a) If the fan IS NOT running continue to part b. If the fan IS RUNNING but, after fifty (50) seconds, the
screen REMAINS BLACK proceed to
page 598
, procedure number
b) Press the Power button for two (2) seconds! If the fan IS NOT running perform parts
c and d. If the fan is running but, after fifty (50) seconds, the screen REMAINS
BLACK proceed to
page 598
, procedure number
c) Plug into a
KNOWN GOOD
GFI power outlet.
d) Press the Power button for two (2) seconds! If the machine still DOES NOT turn on (fan NEVER
runs), ENSURING the GFI did not ‘blow’, proceed to
page 595
, procedure number
8) IF after being on for fifty (50) seconds the screen REMAINS fully OR partially white and possibly
displays nothing:
Proceed to
, procedure number
9) IF “Failed Sending Data To Actuator Board”:
Proceed to
page 597
, procedure number
10) IF after being on (fan running) for fifty (50) seconds the screen displays ‘weird’ data i.e.
scrambled, weird colors, abnormal lines etc.:
Proceed to
, procedure number
11) IF after being on (fan running) for fifty (50) seconds the screen still says “Loading
Software….Please Wait”:
See parts a through c below:
a) To avoid pulling cables loose, GENTLY open the card cage.
b) Behind the card cage (Figure below), ENSURE the 24V POWER harness is plugged in properly.
c) Proceed to
page 599
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...