2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 552
OVER- 8.2.0 MAXIMUM PRESSURE IS BETWEEN 18 AND 20 PSI / ISOLATE PRESSURE REGULATOR
This procedure checks Pressure Regulator #61’s ability to maintain pressure over time.
a)
IMPORTANT!
Turn the machine
OFF
!
b) Leave the water ON!
c) NOTE gauge pressure then allow fifteen (15) minutes before continuing!
d) Does pressure increase more than 2 psi over what was noted in part c?
Yes
Pressure increased more than 2 psi! TWO (2) possibilities:
1)
Incoming water pressure is
more than 105 psi OR;
2)
Bad Regulator #61 NOTE: Regulator #61 can be rebuilt (Rebuild
kit, P/N 190934).
No
Pressure did NOT increase over time. Pressure regulator #61 is okay. The overflow
problem should be solved!