2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 222
T- 1.0.8 ISOLATE TEMPERATURE PROBLEM
a) Watch [
Temperature
] for one (1) minute.
If UNSTABLE
i.e. changes more than 0
.
2° C watch for three
(3) more minutes BEFORE continuing to part b.
If STABLE
i.e. does NOT change more than 0
.
2° C see
part b.
b) FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) unstable AND, at least once, becomes more than 38.9° C:
Proceed
to
page 238
, procedure number
2)
IF (and ONLY if) STABLE between 35.1 and 38.9° C:
Figure
right. is the [
Temperature
] window pale yellow / white OR red?
•
IF (and ONLY if) pale yellow / white!
Proceed to
page 225
, procedure number
•
IF RED!
(page 222)
3)
IF (and ONLY if) unstable and falls to OR remains less than 35.1° C:
Proceed to
page
, procedure number
4)
IF UNSTABLE BUT remains more than or goes to 35.1° C BUT NEVER more than
38.9° C:
Proceed to
page 229
, procedure number
T- 1.0.9 [TEMPERATURE] BETWEEN 35.1 AND 38.9 BUT ITS WINDOW IS STILL RED
a) Figure right, at the distribution board, ENSURE position #4
(“PH-PR”) is VACANT!!
b) ENSURE Cond Cell #7 is plugged into position “X7, COND”!
c) Figure right, the
FOUR-RESISTOR
SET
is required.
d)
Enter Service Mode → Calibrate
Sensors → Temp Sensor
e) Proceed to
STEP #1
(page 222).
STEP #1
a) The screen should say
“Connect a 6.808 K ohm resistor…”
.
b) Per the Figure (above, right), avoiding VACANT position #4, place the 34° C
(6.808 KΩ) plug, from the
FOUR-RESISTOR SET
, into the
2
nd
distribution board position from the left, “MON-NTC”.
c) Is the screen’s
[Pre-Temperature Reference]
between 64 and 76 (Yes or No)?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...