2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 179
CLEAN- 4.4.2 MORE THAN 300 ML / VALVE #24 OPEN / ISOLATE VALVE #25
a) With the dialyzer connector out of the shunt ENSURE the external flow indicator’s ‘bob’ is rising at least
¼ way up in the sight tube!
b) Return the connector to the shunt and close the door.
c) Does the external flow indicator’s ‘bob’ now rise at least ¼ way up?
Yes
‘Bob’ moving now! See procedure number
(page 180).
No ‘Bob’ NOT moving! See parts a AND b below!
a) ENSURE no restrictions in the ‘from dialyzer (red) line’ especially the external filter!
b) If no restrictions,
NOTE
ONLY
VALVE #25
will be checked and proceed to
page 192
LEFT BLANK INTENTIONALLY