2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 429
FIL- 3.6.2 FILACT = 1 / CHECK FOR AIR AT CHAMBER #69
Figure below, watching for two (2) minutes, are air bubbles seen MOVING INTO Air Removal Chamber #69
now?
Yes
Air seen! Proceed to
page 490
SECTION 11 - INDUCED AIR LEAK TESTS
source of the air leak!
No No air seen! Watch
FILACT
again for one (1) minute or until it EVER = 1! THREE (3) possible
scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) FILACT always = 1 (NEVER = 0):
2)
IF (and ONLY if) FILACT now REMAINS = 0:
Proceed to
page 431
, procedure number
3)
IF FILACT cycles between 0 and 1:
A procedure, in different Section, is performed next.
NOTE
this page and procedure number (FIL- 3.6.2) as you may prompted to return here.
Perform parts a THROUGH c below:
a) BEFORE continuing to part a, proceed to
page 527
b) If a leaking Balancing Chamber valve was not located in part a, return to Dialysis
Program and wait until the external flow indicator’s bob moves up and down (i.e.
good Temp and Cond).
c) Call debug screen 1. Watch
FILACT
again for one (1) minute or until it EVER = 1.
TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT EVER = 1:
2) IF FILACT now REMAINS = 0:
Proceed to
page 431
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...