2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 141
SECTION 2 – “NO WATER” ALARM
A) ENSURE the water is on! If attached to a PORTABLE RO ENSURE it is on, alarm free AND NOT in
product divert!
B) ENSURE the incoming water line, between the 2008K/K
2
and the RO, is connected PROPERLY!
C) Call debug screen 0.
D) Ignoring the TOP
Flow Error
window, look at
Valve Error
(2
nd
window
down).
TWO (2) possible scenarios 1) or 2) below:
1)
IF (and ONLY if) Valve Error = 0:
(page 141).
2) IF Valve Error = 1 longer than two (2) seconds consistently:
Refer to the Table below:
If the machine is CURRENTLY in:
Your response
Dialysis Program i.e. connected to
Acid and Bicarb
Proceed to
page 652
A Cleaning / Disinfection Program
(Rinse, Heat Disinfect, etc.)
Proceed to
page 189
, procedure number
NW- 1.0.0 VALVE ERROR = 0
a) Figure below, ENSURE the Vent Tubing is NOT kinked,
INCLUDING
the short segment going back
to the Float!
b) Is the machine attached to a PORTABLE RO? TWO (2) possible scenarios 1) or 2) next page: