2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 267
T- 2.0.5 T5 BETWEEN 20.0 AND 28.0 VOLTS DC
a) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
b) Watching for one (1) minute, if
Flow Error
EVER = 1 OR a “No Water” alarm EVER appears DO NOT
troubleshoot temperature!
c)
CAUTION!
The next measurement is from the Power Logic board’s twenty-pin X2 cable, at
pin 7
.
Pin 9, next door, is connected to 24 volts DC and
touching pins 7 and 9 together WILL
DAMAGE several boards!
To prevent this, TWO (2) precautions:
1) Use the protected red meter lead!
2) Per the Figure below, place the meter lead horizontally onto pin 7
(BOTTOM row,
four pins from the REAR of the machine). 4.0 volts DC or MORE?
Yes
Pin 7
is 4.0 volts OR MORE! Proceed to
page 269
, procedure number
No
Pin 7
is LESS THAN 4.0 volts! See procedure number
(page 268).
Figure 45 - Power Logic Board X2 / Pin 7