2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 273
T- 2.1.8 T5 LESS THAN 2.0 VOLTS
Per the Figure below, measure again at the Power Logic Board’s X2 cable,
pin 6
(TOP row, three pins from
the REAR of the machine). TWO (2) possible scenarios:
1)
IF (and ONLY if) 9.0 volts DC OR MORE:
Proceed to
page 278
2) IF LESS THAN 9.0 volts DC:
(page 274).
LEFT BLANK INTENTIONALLY