2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 76
F- 7.3.2 ISOLATE ‘SPENT SIDE’ FLOW PATHS
a) Valve connectors will be switched to test ‘spent side’ flow paths to
drain. A second bucket and the 1000 ml cylinder is required!
b) Figure right, at the rear of the machine, remove the tubing from the
Drain (
BOTTOM!
) nozzle.
Flow from the nozzle will be measured.
Good flow is more than 800 ml per minute!
c) ENSURING “Select Program” REMAINS up, perform
BOTH
PARTS
3 and 4 EVEN IF PART 3 yields bad flow!
PART 3:
a)
Place the second bucket behind the dr
ain
nozzle.
b) Plug valve #36 into valve #33’s distribution board position, “V33”.
c)
Watch drain flow (if any) for one (1) minute. If good initially but significantly slows down
indicates an intermittent bad valve #32 or #36. If no flow ENSURE valve #36’s
connector is placed properly in position “V33”. If okay see part d!
d) Measure flow, into the cylinder, for one (1) minute.
e)
Return valves #36 AND #33 to their distribution board positions.
f) Record PART 3’s measurement,
THEN perform PART 4!
PART 4:
a)
Refill the bucket that has the dialyzer connector in it! It
MUST NEVER run empty!
b) Plug valve #38 into valve #31’s distribution board position, “V31”.
c) Watch drain flow (if any) for one (1) minute. If good initially but significantly slows down
indicates an intermittent bad valve #34 or #38. If no flow, ENSURE valve #38’s
connector is placed properly in position “V31”. If okay see part d.
d) Measure flow for one (1) minute.
e) Unplug valve #38 THEN analyze BOTH Part 3 AND Part 4’s measurements per
Table 3 – Parts 3 AND 4
PART 3’s
measurement
PART 4’s
measurement
Your Response
Less than 800 ml
Less than 800 ml
(page 78)
Less than 800 ml
More than 800 ml
(page 84)
More than 800 ml
Less than 800 ml
(page 86)
More than 800 ml
More than 800 ml
(page 88)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...