2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 351
Yes
ABic
=
Bic!
(page 351).
No
Watch the Bicarbonate (BIC) pump’s ‘dot’ (Figure right) for TWO (2)
minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) the ‘dot’ EVER turns white:
Proceed to
page 374
, procedure
2)
IF the ‘dot’ REMAINS CONSTANTLY BLUE:
Proceed to
page 397
CO- 5.4.0 ISOLATE ACID PUMP
a) Figure right, ENSURE the Acid (red) AND Bicarbonate
(blue) Connector’s O-rings are present
b) ENSURE the filters is clean AND the adaptor plug is
screwed in tight!
c) Read parts d AND e before performing them.
d) Perform the ACID (CONCENTRATE) PUMP VOLUME CALIBRATION per the CALIBRATION
PROCEDURES booklet.
e) Are you able to accurately enter the burette
volumes and ‘CONFIRM’ them without an
“Operator Error” banner (Figure right) occurring?
Yes
“Operator Error” did NOT occur! See procedure number
(page 351).
No “Operator Error” occurred! Proceed to
page 349
, procedure number
CO- 5.5.0 ISOLATE BICARB PUMP
a) Read parts b AND c before performing them.
b) Perform the BICARBONATE PUMP VOLUME CALIBRATION.
c) Are you able to accurately enter the burette volumes and ‘CONFIRM’ them without an “Operator Error”
banner occurring (Yes or No)?
Yes Operator Error” did NOT occur. Per the Figure
(right), CAREFULLY trace Cond Cell #7’s wire
harness from distribution board position “X7,
COND” to the cell #7 checking for insulation
damage. If no damage, see procedure number
(page 352).
No
“Operator Error” occurred. See procedure
number
(
page 349
).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...