2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 248
T- 1.5.3 SCREEN [TEMPERATURE] IS UNSTABLE
a)
Figure right, insert the 80° C (1.255 KΩ)
plug, from the
FOUR- RESISTOR SET
,
into the
2
nd
distribution board position
from the LEFT, “MON-NTC”.
b) If the plug is placed properly the
screen’s
[Temperature]
window
should read more than 75° C!
c) To prevent damage DO NOT leave the machine. The external meter MUST NEVER reach 95° or more!
d) Once again, watch the screen’s
[Temperature]
for one (1) minute noting its highest and lowest values.
e) Subtract the lowest value seen from the highest. TWO (2) possible scenarios 1) or 2) below:
1)
IF (and ONLY if) the result is more than one (1) i.e. unstable:
Proceed to
page 254
, procedure number
2) IF the result is zero (0) or one (1) i.e. STABLE:
Perform parts A and B below:
A) ENSURING the flow indicator’s ‘bob’ is moving up and down, watch the external meter for one
(1) minute noting its highest and lowest values.
B) Determine stability by subtracting the lowest value seen from the highest. Is the result more than
two (2.0)?
Yes
More than two (2.0)! For future use, NOTE METER STABILITY is more than two (2.0)
THEN see procedure number
(page 249).
No
Less than two (2.0)! For future use, NOTE METER STABILITY is less than two (2.0)
THEN see procedure number
(page 249).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...