2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 422
d)
IMPORTANT!
Allow thirty (30) seconds BEFORE continuing to part e
e) Based on debug screen’s
FILACT
now, TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT ALWAYS = 0:
(page 423).
2) IF FILACT = 1 or cycles between 0 and 1:
Perform parts a THROUGH c below:
a) ENSURE the plug is placed PROPERLY in the Air Sensor’s position! If CBE equipped, using a
flashlight, ENSURE the plug covers ALL CBE board pins especially the TOP pin! If
FILACT
remains = 1 see
part b
!
If
FILACT
now = 0 always see
423).
b) Use a different plug! If
FILACT
remains = 1 see
part c.
If
FILACT
= 0 always see
procedure
(page 423).
c) SIX (6) possible bad components:
1)
CBE board
a
;
2)
Sensor Board
b
;
3)
Sensor Board cable
c
;
4)
Functional Board
b
;
5
) Distribution board;
6)
Motherboard.
a
A)
Swap in a known good CBE board;
B)
Return the resistor plug PROPERLY to the Air
Sensor’s position!;
C)
Return to Dialysis Program (‘Dialysis’ → ‘CONFIRM’);
C)
From debug
screen 1, if
FILACT
now = 0 the previous CBE board may be bad.
b
A)
Swap in a known good board;
B)
For each board, to prevent “Cond Offset” Failure”, place
the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not
defined.
);
C)
Return to Dialysis Program;
D)
From debug screen 1, if
FILACT
now = 0 the
previous board is bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...