2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 494
AIR- 1.0.4 AIR SEEN AT L10 / TRACE UPSTREAM FROM CHAMBER #69
TMP is between 300 and 360 mmHg,
for two (2) minutes,
watch for air
starting at Location #11 (L11).
Table 9 – Air Checks (2)
Location
Description
If air is seen go to:
If NO AIR is seen the leak is from:
L11
Output Blood Leak Sensor
#8 (see
, page
L12 (next row)
Valve #24 / Valve #26 mounting O-Ring(s) (to
DiaSafe
®
Filter Tubing connection
L12
Input Blood Leak Sensor #8
(see
L13 (next row)
Blood Leak Sensor #8
L13
Output Valve #25 (see
L1 (next row)
Dialysate Pressure Transducer #9
L1
OUTPUT (to machine)
side of Dialysate Filter
#73
(see
, page
See (ABOVE) procedure
(
page 492
Valve #25
Figure 78 – Blood Leak Tubing