2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 611
P- G.5.2 SYMPTOM STILL PRESENT / ISOLATE FUNCTIONAL / POWER LOGIC BOARD DEVICES
a)
IMPORTANT! The machine MUST be OFF to prevent damage!
b)
Per the Figure below
, device cables along the TOP EDGE of the Functional AND Power Logic
boards will be unplugged next. To prevent damage when eventually returning them,
RECORD
where
they belong BEFORE continuing to part c!
c) Unplug the cables but
DO NOT
unplug the Power Logic board’s front cable (cable closest to the
screen)!
d)
ENSURE the meter’s black lead remains connected to chassis ground!
e)
IMPORTANT!
Turn the machine on (fan running)! If possible, return to Dialysis Program OR Heat
Disinfect.
f) Re-m12 and / or +5 volts (within range now?). If the symptom was intermittent call debug
screen 1 to watch
5V Est
and / or
12V Est.
Allowing sufficient time, does the symptom you were
troubleshooting reoccur?
Yes
Symptom reoccurs! Proceed to
page 613
, procedure number
No Screen on and /or good voltages! To locate the bad device, perform parts A THROUGH E next
page:
Figure 117 – Card Cage / Functional and Power Logic Board Devices
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...