2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 472
TMP- 6.5.0 “OPERATOR ERROR” OCCURRED / TROUBLESHOOT DIALYSATE PRESSURE
a) ENSURE the transducer protector, at the external meter, is not WET or consider replacing it!
b) Return the Dialysate lines to the shunt and close the door.
c) Turn the machine off and back on.
d) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
e) From the Home screen, select the [Dialysate Flow] window.
f) Set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’!
g)
Read before performing!
Repeat (ABOVE) procedure number
page 470
) but if the
problem reoccurs swap in the listed components (see Component List below), one at a time, in between
returning to procedure number
(page 470) until the problem does not occur indicating the
last component swapped in is the problem!
Component List
1)
Dialysate Pressure Transducer #9
*
;
2)
Sensor Board;
3)
Actuator-Test Board;
4)
Functional Board;
5)
Sensor Board cable;
6)
Distribution board.
*
To LOCATE Transducer #9 refer to
(page 19).
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...