T- 1.3.17 TEMP 6.0 OR LOWER / ISOLATE NTC# 3 CIRCUIT
Call debug screen 5. WITHOUT LOOKING AWAY, watch
TPRE
(top, middle column) for one (1) minute
noting its highest and lowest values. TWO (2) possible scenarios 1) or 2) below:
1)
IF (and ONLY if) TPRE changes more than +/- twenty (20)! TPRE is unstable:
Leaving the 6.04
KΩ plug installed, swap in the listed components (see COMPONENT LIST below), one at a time, with
known good, then return to Dialysis Program and repeat procedure number
(page 240) to
test each new component until
TPRE
does not change more than +/- 20.
COMPONENT LIST
:
1)
Power Logic Board
a
;
2)
Sensor Board
a,b
;
3)
Deaeration motor;
4)
Flow
motor.
a
To LOCATE these boards refer to
(page 9).
b
To prevent “Cond Offset Failure”
place
the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not
defined.
))
2) IF TPRE does NOT change more than +/- 20 (twenty)! TPRE is stable!
Perform parts a THROUGH g
below:
a)
Turn the HEATER Switch ON!
b) To avoid error, read parts c THROUGH h below BEFORE performing them
c) Turn the machine off and replace NTC #3* with a known good . *To LOCATE NTC #3 refer to
(page 218).
d) Figure right, ENSURE the new NTC #3 is plugged
into the
2
nd
distribution board position from the left,
“MON-NTC”!
e) Turn the machine on and return to Dialysis Program
(
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
f) From the Home screen, set [Dialysate Flow] to 500
ml/min and press ‘CONFIRM’.
g) WITHOUT LOOKING AWAY, watch the [
Temperature]
window for up to ten (10) minutes. Does it
EVER exceed 39.0° C even if only once?
Yes
Exceeds 39.0° C! See procedure number
(page 241). The old NTC #3 is
probably okay!
No
NEVER exceeds 39.0° C! The original problem (High Temperature) is no longer occurring!
See (ABOVE) procedure number
(page 222).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...