2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 235
T- 1.0.255 ‘MEASURED’ TEMPERATURE REMAINS BETWEEN 35.1 AND 38.9° C
a) Figure right, if a 274
Ω plug was previously placed at the Cond
Cell’s distribution board position, “X7, COND”, remove it and
RETURN the Cond Cell’s connector.
b) If a
6.04 KΩ plug was previously placed at NTC #3’s position,
“MON-NTC”, remove it then, avoiding the VACANT position on
the right (“PH-PR”), RETURN NTC #3’s connector to the
2
nd
position from the left,
c) Call debug screen 4. Is
TEMP
(lower left) 4.0 or
more?
Yes
TEMP
is 4.0 OR more! See procedure number
(page 235).
No
TEMP
is less than 4.0. NTC #3* is plugged in incorrectly OR is bad. *To LOCATE NTC #3 refer
(page 218).
T- 1.0.26 TEMP IS 4.0 OR MORE / VERIFY TEMPERATURE DISPLAY
From the Home screen, the [
Temperature
] window is, TWO (2) possible scenarios:
1)
IF (and ONLY if) less than 35.1 OR more than 38.9° C:
Proceed to
page 302
, procedure
2)
IF between 35.1 and 38.9° C:
Is the external indicator’s ‘bob’ moving?
Yes
‘Bob’ moving! See procedure number
(page 235)
No
‘Bob’ is NOT moving! Return to (ABOVE) procedure number
page 225
T- 1.0.27 ‘BOB’ IS MOVING UP AND DOWN
Is the machine’s [
Temperature]
window /- 0.3 of the meter’s reading?
Yes
Within
+/- 0.3° C of each other:
(page 236).
No IS NOT /- 0.3° C of each other:
Proceed to
page 302
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...