2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 155
SECTION 3 - FLOW ERRORS IN A CLEANING / DISINFECTION PROGRAM
NOTE:
If you placed the machine in RINSE because a Flow Error was occurring in Dialysis Program return
to Dialysis Program and proceed to
page 20
SECTION 1 - FLOW ERRORS IN DIALYSIS PROGRAM
These procedures troubleshoot Flow Errors that occur in Heat Disinfect, Rinse, etc. Alarm banners include
“Flow Error”, “Flow Inlet Error” and “Flow Recirc Error 1”.
A) Press ‘Escape’ then ‘CONFIRM’ to “Interrupt” the
Program (Figure right).
B)
Figure below
,
did tubing ‘blow off’ a UF Check Valve OR was there an active leak prior to
Interrupting?
Yes
Tubing ‘blew’ off OR an active leak! See procedure number
(page 155).
No Tubing did NOT blow off AND no active leaks! See parts a AND b below:
a)
If the Flow Error was occurring in Heat Disinfect return to it. If not
place the machine in RINSE!
b) Proceed to
page 157
, procedure number
CLEAN- 1.0.0 UF CHECK VALVE TUBING ‘BLOWING OFF’ OR LEAK
a) Reattach ‘blown off’ tubing and dry the area.
b) Per the Figure above, ENSURE the tubing to and from UF Check Valve #63 is NOT kinked!
Parts c through f next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...