2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 315
T- 8.0.0 ISOLATE NTC #2
This procedure checks NTC #2
resistance (Ω) response. An accurate thermometer is required.
If one is not
available replace NTC #2 (see
known good
then return to (ABOVE) procedure
(page 305).
a)
IMPORTANT! Turn the machine OFF!
b) Per the Figure below, unplug NTC #2’s connector from the
1
st
position from the left, labeled “CON-NTC”.
Figure 54 – Hydraulics Rear View – NTC #2
c)
CAREFULLY
(
thin wires!
) open NTC #2’s female distribution board connector.
d) Referring to the Figure above, remove NTC #2 from the hydrochamber to expose its ‘probe end’. Do
NOT lose the O-ring!
e) If (and ONLY if) NTC #2 appears ‘rusty’ replace it then return to (ABOVE)
(page 305). If NOT rusty continue to part f.
f)
Clean with paper towel.
Do not touch it from this point forward to avoid
fouling it!
g) Fill a Styrofoam cup with water, accurately measured, between 36 and 38
o
C!
h) Per the Figure (right), drop NTC #2’s ’probe end’ into the cup.
i)
Per the Figure right, measure RESISTANCE (
Ω) between the
connector’s green and white wires. Referring to
(page 316),
between the minimum and maximum (
KΩ
) range?
Yes
Return to (ABOVE) procedure number
No Replace NTC #2 with a tested known good then return to
(page 305).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...