2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 554
LEAKING- 1.0.0 UF TUBING BLOWING OFF
a) ENSURE the tubing to and from UF Check Valve #63 is NOT restricted!
b) In the next procedure, but NOT YET, the machine will be placed into RINSE
HOWEVER,
if a leak occurs
DO NOT
turn the machine off but instead, “Interrupt” Rinse by pressing the ‘Escape’ key then
‘CONFIRM’ twice!
(page 554).
LEAKING- 1.0.2 ISOLATE RESTRICTION
a) Following there should be:
•
No leaks from the Fluid Sample port (Figure right) ONLY if UF Check Valve
#64 is okay!
•
‘Squirts’ into the bucket ONLY if UF Check Valve #63 is okay!
b) Place the acid and bicarbonate connectors into their rinse ports.
c) Turn the machine on
THEN go
to RINSE PROGRAM!
d) Allow up to four (4) minutes
OR
until if a leak occurs. TWO (2) possible scenarios:
1)
IF (and ONLY if) NO leaks!
Unless UF Check Valve #63’s tubing ‘blew off’ whatever was causing
the leak is no longer present!
2)
IF a leak occurs!
With the “Interrupted” banner up perform
parts AND b below:
a) If (and ONLY if) UF Check Valve’s tubing ‘blew off’ it is oriented incorrectly.
(page 554).
LEAKING- 1.0.4 LEAK OCCURS
a) Call debug screen 0 (Figure
right). If debug does not appear
press ‘Escape’ then call screen
0.
b) Ignoring
Flow Error
(the TOP
window) look at
Valve Error
the
2
nd
window down!
1)
IF (and ONLY if) Valve Error = 1 LONGER THAN two (2) seconds:
Proceed to
page 189
2)
IF Valve Error = 0 OR ‘blinks to 1’ for less than one (1) second:
Turn the machine OFF then
proceed to
page 562
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...