2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 162
CLEAN- 1.2.4 DEAP = 75 OR LESS / ISOLATE DEAERATION MOTOR
a)
If (and ONLY if) the problem was
occurring in Heat Disinfect return to it.
If not stay in RINSE!
b) Figure right, is the Deaeration Motor shaft rotating
COUNTERCLOCKWISE (CCW)?
Yes Rotating CCW! See procedure
(page 162).
No IS NOT rotating CCW! See parts a
THROUGH c below:
a) ENSURE HEAT DISINFECT OR
RINSE is running!
b) Turn the Heater Breaker Switch OFF!
c)
NOTING ONLY
the
DEAERATION MOTOR
will be checked proceed to
page 130
CLEAN- 1.2.5 DEAERATION MOTOR ROTATING CCW / ISOLATE MOTOR STABILITY
Using the handle end of a screwdriver, push HARD on and release the DEAERATION MOTOR SHAFT
several times. Can you make it stop rotating and REMAIN stopped?
Yes
If ABSOLUTELY SURE the motor stops! TWO (2) possible bad components:
1)
Bad
Deaeration Motor (likely bad brushes) OR;
2)
Bad Deaeration Pump head.
No
The motor DOES NOT stop! See procedure number
(page 163).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...