2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 402
EOS- 5.0.0 BOTH PUMP SYMBOLS ARE PINK
a)
IMPORTANT! Turn the machine OFF to prevent damage!
b) Per the Figure below, ENSURE BOTH pumps are installed with their output nozzles (red or blue)
towards the top and the ribbon cables extending from the side.
If not, this may be the problem!
c) Swap in a known good Actuator-Test Board* *To LOCATE the board refer to
d) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
e) Call debug screen 0. If debug does not appear press ‘Escape’ then call screen 0.
f) Allow thirty (30) seconds THEN WITHOUT LOOKING AWAY, watch BOTH pump symbols for two (2)
minutes. Are BOTH still pink?
Yes
Both pink! See procedure number
(page 402).
No Neither pump pink! The previous Actuator-Test Board is bad.
EOS- 5.0.1 ISOLATE THE PUMPS
a)
IMPORTANT! Turn the machine off to prevent damage!
b) Per the Figure above, plug in a known good pump into distribution board position “P16, “Conc-P” (acid
pump)
*
OR “P16, Bic-P” (bicarbonate pump)
*
. * Your choice.
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press ’Escape’ then call screen 0.
Part e next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...