2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 298
T- 6.0.0 UNSTABLE TEMPERATURE / ISOLATE NTC’S
a) Per the Figure below, ENSURE both NTCs’ are plugged into their distribution board positions. They may
have been unplugged in a previous procedure!
b) Trace EACH CABLE, from the distribution board to the NTC, to once
again
THOROUGHLY
check for
insulation damage!
Insulation damage causes instability!
(page 298).
T- 6.1.0 ISOLATE FLOW ERROR
a) From the Home screen, ENSURE [Dialysate Flow] is set at the rate where the temperature instability is
occurring.
b) Call debug screen 0. WITHOUT LOOKING AWAY, watch
Flow Error
for three (3) minutes or until if it
EVER = 1. TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1:
Proceed to
page 20
2) IF Flow Error = 0:
(page 299)