2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 375
CO- 9.8.0 VERIFY BALANCING CHAMBER VOLUME
Figure below, the
B
alancing
C
hamber(s)
V
olume (BCV) MAY BE factory recorded, a three-digit number
(example
61.3
), written on the top of the Balancing Chambers (Yes or No)?
Yes
BCV is factory recorded!
See procedure number
No
BCV
not recorded! If (and ONLY
if) troubleshooting bad lab values
proceed to
page 376
. All
other problems see procedure
number
CO- 9.8.11 BALANCING CHAMBER VOLUME IS FACTORY RECORDED
Call debug screen 3.
BCV
(upper right) is the stored
B
alancing
C
hamber
V
olume, of
BOTH chambers (example: 613 = 61.3 ml). Does
BCV
= the factory value?
Yes
BCV
= the factory value! If (and ONLY if) troubleshooting bad lab values proceed to
page 377
. All other problems see procedure number
No
BCV
does NOT = the factory value! See parts a THROUGH e below:
a) Perform the BALANCING CHAMBER VOLUME CALIBRATION,
BUT instead of
measuring from the drain enter the factory volume!
b) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) Call debug screen 3. If the calibration took
BCV
now = the factory value.
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’
e) Allow six (6) minutes. Does the machine’s [
Conductivity
] window stabilize to /- 0.4
mS of TCD?
Yes
If (and ONLY if) troubleshooting bad lab values proceed to
page 377
. All other problems see procedure number
(page 375).
No See (ABOVE) procedure number
page 370
CO- 9.8.2 VERIFY ORIGINAL PROBLEM
Did you originally start troubleshooting On Line Clearance (OLC) problems?
Yes
Troubleshooting OLC problems! Proceed to
page 378
, procedure number
No A Conductivity problem is not indicated at this time. DO NOT continue!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...