2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 133
MOTORS- 4.0.0 MOTOR NOT ROTATING
Hit the
NOTED
motor
HARD
with a screwdriver handle. Does it start rotating?
Yes Motor rotating! Proceed to
page 138
, procedure number
No Motor does NOT start rotating! Perform parts a THROUGH e below:
a)
To prevent damage TURN THE MACHINE OFF!
b) Figure below, remove the distribution board cover.
c) TWO (2) checks:
CHECK #1:
ENSURE the pump is plugged in PROPERLY!
CHECK #2:
ENSURE the Actuator Ribbon Cable is plugged in securely!
d) Turn the machine ON.
e) Depending on what Program the machine was in when you ORIGINALLY started
troubleshooting:
1)
IF a Cleaning Program (Heat Disinfect, Rinse, etc.):
Place the machine in RINSE then
(page 134).
2)
IF Dialysis Program:
Perform parts A through C below:
A)
Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
B)
From the Home screen, set [Dialysate Flow] to 800 ml/min.
C)
See procedure number