2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 556
5)
IF from RINSE PORT tubing:
Secure tubing connections with tie wraps.
6)
(page 557),
IF from Regulator #61 OR Valve #41 (not including the Heat
Exchanger):
A)
Ensure incoming water pressure is not exceeding 100 psi;
B)
Check Inlet Water
Pressure Regulator per the CALIBRATION PROCEDURES booklet;
C)
Check tubing connections tie
wraps or replace tubing;
4)
If Valve #41 is leaking externally the tubing connection, O-ring, or the valve
body may be cracked.
7)
(page 556),
IF from Valve #39: A)
Check the tubing connection;
B)
Possible bad
O-ring between the valve and the Hydrochamber;
C)
The valve body may be cracked.
8)
IF from the Acid or Bic pump reinforced (output) tubing:
Secure tubing connections with tie wraps.
9)
(page 556),
IF between Deaeration Pump #20 and the Hydrochamber including
Mixing Chambers #82:
Secure
tubing connections with tie wraps.
10)
(page 557),
IF between BOTTOM BALANCING CHAMBER VALVES, #35 / #37 and
the Hydrochamber:
A)
Secure tubing connections with tie wraps
;
B)
Possible bad O-ring between the
valve and the balancing chamber;
C)
A valve body may be cracked.
11)
(page 557),
IF between TOP BALANCING CHAMBER VALVES #32 / #34 and the
HEAT EXCHANGER: A)
Secure
tubing connections;
B)
Possible bad O-ring between the valve and
the balancing chamber;
C)
A valve body may be cracked.
Figure 105 A – Various Hydraulic Views
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...