2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 527
SECTION 18A - DIAGNOSTIC VALVE LEAK TESTS
Perform these tests ONLY as directed previously. Arbitrary performance may yield erroneous results!
ONLY the Balancing Chamber Valves, #31 through #38, will be tested!
Do NOT run “auto” which tests
all valves!
A. AUTOMATED LEAK TEST PREPARATION
a) Return both dialyzer connectors to the shunt and close the door!
b)
IMPORTANT!
Return the concentrate connectors to the rinse ports and go to RINSE PROGRAM!!
c) Using a flashlight, THROUGHLY check for and repair ANY leak located!
d) E
nter Service Mode → Diagnostics → Valve Leak Test. When the
[Test Status]
data box says “Ready”
proceed to
B. PERFORM IOS
TEST
The
IOS
(
I
ntegrity
O
f
S
ystem)
isolates Chamber #69 and its tubing for leaks AND an
IOS
failure may cause
valves to fail falsely!
IOS must pass BEFORE testing the valves!
a) Select the screen’s
[Valve Number]
data box.
b) K machine use the up/down arrow keys until
[Valve Number]
= “IOS”; K
2
machine use the +/- keys until
[Valve Number]
= “IOS”.
c) Press ‘CONFIRM’ and allow “Prep” and “Testing” to complete then read the
[TEST STATUS]
box. TWO
(2) possible scenarios:
1)
IF “Failed”:
REPEAT the
IOS
test. If (and ONLY if) it passes see Scenario #2 below. If it fails
again turn the machine off then see
(page 528).
2) IF “Passed”:
Perform parts a THROUGH d below until
ALL EIGHT
Balancing Chamber Valves,
#31 through #38, have been tested:
a) Select the
[Valve Number]
box THEN: K machine use the up/down arrow keys until
[Valve
Number] =
a Balancing Chamber Valve
*
; K
2
machine use the +/- keys select until
[Valve
Number]
= a Balancing Chamber Valve
*
.
*
The ‘fresh’ valves, #31, #33, #35 and #37 are tested first followed by the ‘spent’ valves, #32,
#34, #36 and #38.
b) Press ‘CONFIRM’ and allow “Prep” and “Testing” to complete.
c) If a valve passes ONCE consider it “passed” and continue to part d. If a valve fails repeat parts a
and b on THAT valve. If it fails again consider it “failed” THEN
NOTE
the valve number(s) AND
continue to part d.
d) Repeat parts a THROUGH d until
ALL EIGHT
balancing chamber valves have been tested then
refer to
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...