2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 98
F- 9.0.41 LESS THAN 800 ML TO VALVE #26
a) ENSURE no restrictions or leaks between the DiaSafe
®
filter and the Balancing Chambers.
b) ENSURE no restrictions between the DiaSafe
®
filter and Valve #26.
c)
Figure below
, at the rear of the DiaSafe
®
filter housing, remove the tubing from Valve #28 (V28) and
direct it into the bucket.
d)
Return to Calibrate Hydraulics → Flow Pressure.
e) Press ‘CONFIRM’ then ENSURE the Flow Motor is rotating!
f) Direct the tubing into the graduated cylinder for one (1) minute.
g) Press ‘CONFIRM’ to return to “Calibrate Hydraulics”.
h) More than 800 ml collected?
Yes
More than 800 ml collected! The DiaSafe
®
filter is restricted!
No
Less than 800 ml collected! This indicates a restricted ‘Fresh Side’ Balancing Chamber valve.
See parts THROUGH c below:
a) Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program (“
Select Program” → ‘Dialysis’ →
‘CONFIRM’).
c) Proceed to
page 62
procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...