![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 394](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635394.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 390
CO- 10.0.50 OLC TEST STARTED / OLC SELF TEST DESCRIPTION / TROUBLESHOOTING
a) Call debug screen 5. When
OLC Status
= 1, Conductivity Pre-Dialyzer (
CPRE
)
followed a minute or so later by Conductivity Post-Dialyzer (
CPOS
) SHOULD
gradually increase to more than 15000 (15.000 mS)! The initial instability causes
Pre Stbl Ct
AND
Post Stbl Ct
to increase to respectively 90 and 70.
b) After no more than four (4) minutes, if
CPRE
stabilizes at more than 15.0 mS,
Pre Dev val
falls. If it
falls to and remains less than 135
*
then
Pre Stbl Ct
counts down.
c) After no more than (5) five minutes, if
CPOS
stabilizes at more than 15.0 mS,
Post Dev val
falls. If it
falls to and remains less than 70
*
then
Post Stbl Ct
counts down.
*
If a BLOOD ALARM OR a red banner Conductivity alarm occurs the counters stop! In this event, see
NOTES 1 and 2 below!
NOTE 1:
Resetting Blood Alarms is okay but multiple resets may cause OLC to cancel. Having to
perform multiple Blood alarm RESETS indicates a problem unrelated to Conductivity!
NOTE 2: IF a red banner Conductivity alarm occurs:
A)
Turn the machine OFF;
B)
ENSURE
the acid and bicarb is GOOD;
C)
Return to Dialysis Program with [Dialysate Flow] set to
500 ml/min;
D)
Allow six (6) minutes;
E)
Return to (ABOVE) procedure number
(page 386). If (and ONLY if) you return to here, because a Conductivity
alarm reoccurs, see (ABOVE) procedure number
(page 354).
d) When (and if) BOTH
Pre Stbl Ct
AND
Post Stbl Ct
= 0 (Conductivity is stable),
OLC Status
= 2.
e)
CPRE
followed a minute or so later
CPOS
SHOULD gradually fall until less than 14.0 mS. The initial
instability causes
Pre Stbl Ct AND Post Stbl Ct
to increase. After a few minutes, if
CPRE
stabilizes
Pre Stbl Ct
counts down. A couple of minutes later, if
CPOS
stabilizes,
Post Stbl Ct
counts down.
f) When
Pre Stbl Ct
AND
Post Stbl = 0
then
OLC Status
= 0 indicating the OLC test is complete!
g) THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) “OLC Test Passed”: Proceed
to
page 395
, procedure number
2) IF (and ONLY if) “OLC Test Failed” AND a Conductivity alarm did NOT occur during the
test:
(page 391).
3)
IF “OLC Test Cancelled”:
Call debug screen 6 to see the
OLC Cancels 1 – 6
data boxes. If
code “1” and / or “2” appear in any of the boxes see procedure number
391). Codes “3” through “7” refer to
(page 395) for their meaning.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...