2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 289
T- 4.0.0 TEMPERATURE EXCEEDS 39.0° C / ISOLATE TEMPERATURE CONTROL
a) So as NOT to pull cables loose,
GENTLY
open the card cage.
b)
Set your CALIBRATED volt meter to DC volts (V
DC
)!
c)
Attach the meter’s black lead to chassis ground (see
d)
CAUTION!
A signal will be measured from a pin that is VERY close to others and touching pins
together with a standard meter lead WILL CAUSE DAMAGE! As directed below, make your RED meter
lead PROTECTED!
DO NOT CONTINUE UNTIL YOU HAVE DONE THIS!
e) Per the Figure below, at the top of the Power Logic Board, closest to the screen, locate its 20-pin X2
cable.
f) Measure at
pin 6
(TOP row, 3 pins from the rear of machine). More than 1.0 volts DC?
Yes
MORE THAN than 1.0 volt DC! See procedure number
(page 290).
No
LESS than 1.0 volt DC! Proceed to
page 292
, procedure number
Figure 50 – Power Logic Board X2 / Pin 6
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...